Phaco
System Specifications
|
|
Mode
|
Specification
|
Function
|
Specification
|
|
IRR
|
Irrigation
only, gravity feed
|
Panel
|
Mode
enabled at fully preset values
|
|
I/A
|
Adjustable
aspiration rate
(Rate): 0 to 40 cc/min
Adjustable vacuum limit (I/A VAC):
0 to 500 mmHG |
Surgeon
|
Footpedal
control of aspiration rate in I/A,
VIT
Footpedal control of U/S power in
U/S modes
|
|
U/S STD
|
Adjustable
U/S power
(U/S Rate): 0 to 100%, 40 kHz
Adjustable aspiration rate
(Rate): 0 to 40 cc/min
Adjustable vacuum limit
(U/S VAC): 0 to 200 mmHg
|
Cautery
|
0 to 100%,
10 W into 600 ohms, 50 W max
|
|
U/S Pulse
|
As
above except U/S power is
pulsed at 10 pulses/sec
|
Reflux
|
Rate
of 25 cc/min
|
|
VIT
|
Adjustable
cut rate
(VIT): 60 to 600 cuts/min, single
cut
Adjustable aspiration rate
(Rate): 0 to 40 cc/min
Adjustable vacuum limit
(U/S VAC): 0 to 200 mmHg
|
Audio
|
Audible
tone, frequency increases with vacuum
|
|
Electrical
|
100
to 240 VAC, 50 to 60 Hz
|
Display
|
Specification
|
|
|
Console:
|
Timer
|
Accumulated
U/S time w/rese
|
|
|
Depth:
32 cm (13 in)
Height: 11 cm (4.4 in.)
|
VAC
|
Preset
vacuum with footpedal at rest
Actual vacuum with footpedal depressed
|
|
|
Width:
40 cm (16 in)
Weight: 8.4 kg (18.5 lbs)
|
U/S Power
|
Preset
U/S power with footpedal at rest
Actual U/S power with footpedal
|